The formula
Concentration (ng/µL) = (A260 − A320) × extinction factor × dilution factor ÷ path length (cm)
Factors: dsDNA 50, ssDNA/oligo 33, RNA 40 ng/µL per absorbance unit
Purity: A260/A280 ≈ 1.8 for DNA, ≈ 2.0 for RNA; A260/A230 ideally 2.0–2.2
Molar mass: dsDNA = bp × 617.96 + 36.04 g/mol; RNA = nt × 320.5 + 159
nM = (ng/µL) ÷ molar mass × 10⁶; copies/µL = ng/µL × 10⁻⁹ ÷ molar mass × 6.022 × 10²³
Worked example
- A plasmid sample diluted 1 : 10 reads A260 = 0.75 in a 10 mm cuvette:
- Concentration = 0.75 × 50 × 10 = 375 ng/µL; 50 µL contains 18.75 µg.
- A260/A280 = 0.75 ÷ 0.40 = 1.88 (pure DNA); A260/A230 = 2.08 (clean).
- A 3,000 bp plasmid has a molar mass of 1,853,916 g/mol, so 375 ng/µL is 202 nM, or about 1.2 × 10¹¹ copies per µL.
Measuring nucleic acids by UV absorbance
DNA and RNA absorb ultraviolet light strongly at 260 nm because of the aromatic rings in their bases. By the Beer–Lambert law, absorbance is proportional to concentration, so a spectrophotometer reading at 260 nm gives a quick measure of how much nucleic acid a sample contains. Instruments such as the NanoDrop measure tiny 1–2 µL drops and normalise to a 10 mm path length automatically. With a standard cuvette, you usually dilute the sample so the A260 falls between about 0.1 and 1.0, where readings are most accurate.
Conversion factors
An A260 of 1.0 in a 1 cm path corresponds to about 50 ng/µL of double-stranded DNA, 40 ng/µL of RNA and 33 ng/µL of single-stranded DNA. These are average values: single-stranded molecules absorb more per mass because their bases aren't stacked in a double helix. For short oligonucleotides, the exact extinction coefficient depends on the sequence, so oligo suppliers often provide a specific factor that you can enter with the custom option. Remember to multiply by the dilution factor if you diluted the sample before measuring.
Purity ratios
The A260/A280 ratio checks for protein contamination, since proteins absorb at 280 nm. Pure DNA gives about 1.8 and pure RNA about 2.0; lower values suggest protein or phenol carried over from extraction. The A260/A230 ratio checks for contaminants such as guanidine salts, EDTA, carbohydrates and phenol, which absorb near 230 nm; values of 2.0–2.2 indicate a clean sample. An A320 reading measures background from turbidity or bubbles and is subtracted from the other readings. Low ratios don't always ruin experiments, but they can inhibit enzymes in PCR and sequencing.
From ng/µL to molarity and copy number
Many protocols need molar amounts rather than mass: ligation reactions, sequencing libraries, qPCR standards and transfections often specify nanomolar concentrations or numbers of copies. To convert, divide the mass concentration by the molecule's molar mass, which depends on its length. A base pair of double-stranded DNA averages about 618 g/mol, so a 3 kb plasmid is about 1.85 MDa. Copy number per µL follows from Avogadro's number and is used for qPCR standard curves and estimating genome copies.
Limitations of absorbance
UV absorbance can't distinguish DNA from RNA or free nucleotides, so RNA contamination inflates apparent DNA concentration, and degraded samples still absorb. For low concentrations, below about 2–5 ng/µL, readings become unreliable. Fluorescent dye assays, such as Qubit and PicoGreen, bind specifically to double-stranded DNA or RNA and are more accurate for low-concentration and sequencing samples. Gel or capillary electrophoresis checks integrity and size. Use absorbance for a quick estimate and purity check, and dye-based methods when accuracy matters.
Storing DNA and RNA samples
Store purified DNA in a slightly alkaline buffer such as TE (10 mM Tris, 1 mM EDTA, pH 8) or nuclease-free water at −20 °C, and avoid repeated freeze–thaw cycles, which shear long molecules. RNA is far less stable and is easily degraded by RNases on skin and surfaces, so keep it on ice during work and store it at −80 °C. Mix thawed samples gently before measuring, because concentrated DNA can settle and give misleading readings.
Related lab calculators
After measuring concentration, use the dilution calculator to prepare working dilutions at a target ng/µL or nM, and the PCR primer Tm calculator to set up amplification. For buffer recipes, the molarity calculator gives the mass of each component, and the molar mass calculator handles any chemical formula.